3 connection of a printer – Metrohm 799 GPT Titrino User Manual
Page 213

5.1 Setting up and connecting the instruments
799 GPT Titrino
209
5.1.3 Connection of a printer
A variety of printers can be connected to the RS232 interface of the
Titrino. If you connect a printer other than one of those mentioned
below, ensure that the Epson mode is emulated or that it uses the in-
ternational character set following the IBM Standard Table 437 and
IBM-compatible graphics control characters.
If a balance is connected at the same COM of the Titrino as a
printer, you need the 6.2125.010 + 6.2125.030 Adapters.
Printer
Cable
Settings on Titrino
Settings on Printer
Seiko
DPU-414
6.2134.110 baud
rate:
9600
data bit:
8
stop bit:
1
parity: none
handshake: HWs
send to:
Seiko
none
Custom
DP40-S4N
6.2134.110 baud
rate:
9600
data bit:
8
stop bit:
1
parity: none
handshake: HWs
send to:
Citizen
none, pre-set on Metrohm-version
IDP-560 EMULATION
FONT MAP =GERMANY
PRINT=REVERSE
LITTLE
CR CODE=VOID
CR AFTER B :FULL=VOID
CR ON b. EMPTY=VALID
BUFFER 1K BYTE
BAUD RATE=9600
PROTOCOL=8,N,1
FLOW CONTROL CTS-RTS
Citizen
iDP562 RS
6.2134.050 baud
rate:
9600
data bit:
8
stop bit:
1
parity none
handshake: HWs
send to:
Citizen
Epson LX-
300+
6.2134.050 as above but
send to:
Epson
see printer manual
HP Desk Jet
with serial
interface
6.2134.050 baud
rate:
9600
data bit:
8
stop bit:
1
parity none
handshake: HWs
send to:
HP
HP Desk Jet
with parallel
interface
2.145.0330
RS232/
Parallel
Converter
baud rate:
9600
data bit:
8
stop bit:
1
parity none
handshake: HWs
send to:
HP
see printer manual
ON
1
2
3
4
5
6
7
8
9
10
SSW1
1
2
3
4
5
6
7
8
1
2
3
4
5
6
7
8
B:
A4 paper
A: