2.12 TIP, Titration procedure

799 GPT Titrino

87

Sequence of TIP
As there is no preset sequence of TIP, in what follows the procedure is illustrated by a

sequence that contains all available commands.

The start delay time is waited off.

The sample identifications and the sample size

are requested.

Stirrer is switched on. The stirrer is not operated

automatically in submethods of TIP. At the end of

TIP, however, the stirrer is switched off (with stir-

rer control ON).

Submethods are processed according to their

parameters. They run to completion, including

calculations and data output (e.g. curves). The

determination data of submethods are overwrit-

ten in the next sequence step of TIP. Those val-

ues which are needed for higher-level calcula-

tions must therefore be assigned to temporary

variables C7X.

The pause is waited off.

Outputs on the socket "Remote" can be set.

A message can be written in the display. The se-

quence is held until it is continued manually (with

, or ).

The titration burets can be prepared.

TIP higher-level calculations are carried out.

Data output in TIP (TIP contains no more deter-

mination data, i.e. curves must be put out within

the submethods).

(Start delay)

(Request ident.)

(Request smpl size)

Submethod

Calculations

Data output

Switching ON stirrer

Set output

TIP calculations

Pause

Info

C7X

TIP data output

Prep