beautypg.com

4 connection of a balance – Metrohm 799 GPT Titrino User Manual

Page 214

background image

5.1. Setting up and connecting the instruments

799 GPT Titrino

210

5.1.4 Connection of a balance

The following balances can be connected to the RS232 output of the

Titrino:

Balance Cable

Sartorius MP8, MC1

6.2134.060

Shimadzu BX, BW

6.2125.080+6.2125.010

Settings on Titrino: balance SARTORIUS

Balance: elimiter CR+LF

Ohaus Voyager, Explorer,

Analytical Plus

from Ohaus: AS017-09 (Ohaus parts number)

Settings on Titrino: balance SARTORIUS

Balance: SET BALANCE

INTERFACE

BAUD RATE 9600

DATA BITS 8

PARITY none

STOP BITS 1

Mettler AB, AG (LC-RS25)

in the scope of delivery of the balance

Mettler AM, PM

6.2146.020 + 6.2125.010

additionally from Mettler:

ME 47473 Adapter and ME 42500 hand switch or ME

46278 foot switch

Mettler interface 016

Cable in scope of delivery of interface 016: red lead

to pin 3, white lead to pin 7 of the 25-pin connector

+ 6.2125.010 25-pole/9-pole adapter

Mettler interface 011 or 012

6.2125.020 + 6.2125.010

Mettler AT

6.2146.020 + 6.2125.010

Mettler PG, AB-S

6.2134.110

Mettler AX, MX, UMX

6.2134.120

AND Models ER-60, 120, 180, 182

Models FR-200, 300

Models FX-200, 300, 320

with RS232 interface (OP-03)

6.2125.020 + 6.2125.010

Precisa, balances with RS232C-

interface

6.2125.080 + 6.2125.010

The balance type must be preselected at the Titrino with the

key.

The weight is transferred as a number with up to 6 digits, sign and

decimal point. Units and control characters sent by the balance are

not transmitted.

With the aid of a special input unit supplied by the balance manufac-

turer, in addition to the weight identifications and methods can be