4.1 Error and special messages

799 GPT Titrino

189

no method

The method required by the sample data from the silo memory

or in a TIP sequence is not available in the method memory.

Exit: .

no new com.var.

The common variable could not be assigned as the result or

the mean value could not be calculated. The old value remains

in force.

no new mean

No new mean value has been calculated as at least one

quantity stipulated for mean value calculations could not be

calculated.

no new silo result

No new silo result C24 or C25 could be stored as the assigned

quantity could not be calculated.

no new temp.var.

No new temporary variable C7X could be stored as the

assigned quantity could not be calculated.

no oven param.

The oven could not be found at the given COM.

Remedy: connect the oven to the given RS-interface of the Ti-

trino or set the following in your method under ,

>preselections, "Oven: no".

no sequence

No sequence is defined in TIP.

Exit: and define sequence.

no titration data

No curve can be printed as no data are available. With the

curves measured value vs. time or temperature vs. time, the

corresponding monitoring function must be on for the

recording of measured points.

not possible

Writing functions on the card 6.6029.XXX are not possible or

the function "empty" cannot be executed with a Dosimat.

not valid

A value is not available.

outside

The set fix volume, a fix EP or fix time is outside the measuring

range.

overrange

The measuring range of ±2 V has been exceeded. Overrange

replaces the corresponding measured value (pH, U, I or

temperature). If a measured value is in overrange (primary or

secondary measured value), the other (secondary or primary

measured value) can also be unstable.

prep.internal D0!

Warning that the preparation for the internal buret D0 has to be

executed (the warning interval is run-off).

Exit: Execute preparation with or ignore warning

with .

prep.external DX!

Warning that the preparation for the external buret DX has to

be executed (the warning interval is run-off).

Exit: Execute preparation with or ignore warning

with .

prep: manual stop

The preparation or emptying of the buret was stopped with

.

rate lim

Message in the measuring point list. A limit value of the rate

monitoring has been violated.

rate out of limit

The determination has been stopped as a limit value of the

rate monitoring has been violated.

rate too high

With DOS the calculated rate is too high. It can not be

achieved with the mounted Exchange Unit.

Exit: and use Exchange Unit with larger buret

volume or change dosing conditions.