beautypg.com

Metrohm 799 GPT Titrino User Manual

Page 195

background image

4.1 Error and special messages

799 GPT Titrino

191

system error 14

No communication between the Titrino and the connected

Remote Box.

Possible causes:

. The Remote Box was connected when the Titrino was

running

. Titrino has a fault.

. Remote Box has a fault.

Remedy: Set under , >peripheral units, "Remote

Box: OFF", switch off Titrino, take away Remote Box and

switch on Titrino. Contact Metrohm service.

temp out of limit

The determination has been stopped as a limit value of the

temperature monitoring has been violated.

temp

Message in the measuring point list: A limit value of the

temperature monitoring has been violated.

time-out PC keyboard

A connected PC keyboard has been used to call up an ad-

dress (e.g. ) and the connection has then been inter-

rupted.

Possible causes:

. Remote Box has a fault.

. PC keyboard has a fault.

Exit: Correct fault and switch Titrino off/on.

TIP terminated

TIP has been terminated.

transmission error

With a Remote Box connected characters are received which

cannot be interpreted.

Possible causes:

. Wrong key combination has been pressed.

. Wrong PC keyboard has been selected.

. The barcode reader supplies garbled characters.

. The Remote Box has a fault.

Exit: Rectify fault and switch Titrino off/on.

validate instrument

Validation interval has elapsed.

Exit: or new start.

wrong card

The card was inserted/removed during an operation with

.

Exit:

wrong sample

With SET, KFT or DOC with preset titration direction the first

measured value is outside the end point.