6.1 Technical specifications
799 GPT Titrino
219
6 Appendix
6.1 Technical specifications
Modes
DET: Dynamic Equivalence Point Titration
MET: Monotonic Equivalence Point Titration
SET: Set End point Titration
KFT: Karl Fisher Titration
STAT: Keeping a measured value constant
DOS: Dosing
DOC: Controlled dosing with measured value gradient
MEAS: Measurement
CAL: pH calibration
TIP: Links commands to titration procedure
Measuring inputs
2 high-impedance measuring inputs for pH, redox and ISE
electrodes.
1 reference input for a separate reference electrode. May
also be used as a differential amplifier.
1 measuring input for polarized electrodes.
1 measuring input for temperature sensor Pt100 or Pt1000.
Measuring range
pH value (pX)
0...±20.00, resolution 0.01
Voltage
0...±2000 mV, resolution 1 mV, error limit 0.1 % fullscale
Current
0...±200
µA, resolution 1 µA
Temperature
–150.0...+450.0 °C, resolution 1 °C
Polarizer Ipol
0...±127
µA, resolution 1 µA
Upol
0...±1270 mV, in steps of 10 mV
Measuring amplifier (at 25 °C and Titrino warmed-up)
Input
resistance
>10
13
Ω
Offset
current
<3 ∗10
–13
A
Deviation of offset voltage
15 µV/K
Dosification
Volume of buret cylinder 1, (2), 5, 10, 20 or 50 ml
Resolution
10 000 steps per buret cylinder
Titrating burets
2 external and 1 internal buret
external: 685 Dosimat or 700 Dosino
Auxiliary burets
2 additional burets: 776 or 765 Dosimat
Materials
Housing
Polybutyleneterephthalate
(PBTP)
Keypad cover
Polycarbonate (PC)