3.2. Remote control commands
799 GPT Titrino
124
*Parameter
Tree part "Parameters for DOS"
├
.
DosPara
Dosing parameters
│ ├
.Type
Type of dosing
volume, time, rate
3.2.2.52.
│ ├
.Volume Volume
dosing
│ │ ├
.Volume
Volume 0...10...99 999.99
ditto
│ │ ├
.DisType
Second dosing criterion
time, rate ditto
│ │ ├
.Rate Rate
0.001...150,
max. ditto
│ │ ├
.Time
Dosing time
1...100...999 999
ditto
│ ├
.Time Time
dosing
│ │ ├
.Time
Dosing time
1...100...999 999
ditto
│ │ ├
.DisType
Second dosing criterion
volume, rate
ditto
│ │ ├
.Rate Rate
0.001...150,
max. ditto
│ │ ├
.Volume
Volume 0...10...99 999.99
ditto
│ ├
.Rate Rate
dosing
│ │ ├
.Rate Rate
0.001...150,
max. ditto
│ │ ├
.StopType
Second dosing criterion
volume, time
ditto
│ │ ├
.Time
Dosing time
1...100...999 999
ditto
│ │ ├
.Volume
Volume 0...10...99 999.99
ditto
│ ├
.Pause
Waiting time after start volume
0...999 999
3.2.2.53.
│ ├
.TDelta
Time interv.for meas.acquisition
1...10...999 999
3.2.2.54.
│ ├
.DosUnit
Selection of the dosing unit
internal D0, external D1,
│ │
external
D2
3.2.2.17.
│ ├
.Temp
Titration temperature
-170.0...25.0...500.0 3.2.2.55.
│ │
├
.
StopCond
Stop conditions
│ ├
.VStop
Stop volume
│ │ ├
.Type
Type of stop volume
abs., rel., OFF 3.2.2.20.
│ │ ├
.V
Volume for absolute stop volume
0...99.99...9999.99 ditto
│ │ ├
.Factor
Factor for relative stop volume
0...±999 999 ditto
│ ├
.FillRate Filling
rate
0.01...150.0,
max. 3.2.2.23.
│ │
├
.
Statistics
Statistics
│ ├
.Status
Status of statistics calculation
ON, OFF 3.2.2.24.
│ ├
.MeanN
No.of individual determinations
2...20 ditto
│ ├
.ResTab Result
table
│ │ ├
.Select
original,delete n,delete all
ditto
│ │ ├
.DelN
Deletion of individual results
1...20 ditto
│ │
├
.
Monitoring
Monitoring
│ ├
.MeasVal
Monitoring of measured values
│ │ ├
.MeasMode
Measured value
pH, U, Ipol, Upol, OFF 3.2.2.56.
│ │ ├
.MeasInput Measuring
input
1, 2, diff.
ditto
│ │ ├
.PCurrent Polarization
current
0...1...±127 ditto
│ │ ├
.PVoltage Polarization
voltage
0...400...±1270 ditto
│ │ ├
.PolElectrTest
Test for polarized electrodes
ON, OFF ditto
│ │ ├
.pH
Limits for monitoring of pH values
│ │ │ ├
.LowLim Lower
limit -20.00...20.00 ditto
│ │ │ ├
.UpLim Upper
limit
-20.00...20.00 ditto
│ │ ├
.U
Limits for monitoring of U values
│ │ │ ├
.LowLim Lower
limit -2000...2000 ditto
│ │ │ ├
.UpLim Upper
limit
-2000...2000 ditto
│ │ ├
.Ipol
Limits for monitoring of Ipol values
│ │ │
:
as for U
│ │ ├
.U
Limits for monitoring of Upol values
│ │ │
:
as
above
│ │ ├
.Action
Action if out of limits
end, hold, wait, none ditto