4.2.

Diagnosis

799 GPT Titrino

197

• Press .

diagnose press key 0...9

6

Analogue input test

• Press <7>.

analog input test 1...5

6.1

Examination of highly insulated measuring inputs

• Connect ‘Ind I’ measuring input to a voltage calibrator (e.g. Metrohm 773 pH simu-

lator) by means of a highly insulated cable (e.g. 6.2108.060). Set calibrator to 0 V.

• Press <1>.

Input 1

0.0 mV

Tolerance: ± 0.5 mV

• Set the calibrator voltage, on setting 'low ohmic' (with 773 = ~ 0.002 MΩ) to differ-

ent values (e.g. +1500 mV) and compare with the displayed value.

Tolerance (with ± 1500 ÷ 2000 mV) ± 1 mV. Be aware of the calibrator's tolerance.

• Set simulator to high ohmic condition (with 773 = 1000 MΩ).

The displayed reading may vary slightly only (with 1500 mV ≤ 1 mV)

• Press .

analog input test 1...5

• Disconnect simulator from ‘Ind I’ input and connect to ‘Ind II’ input.

• Press <2>.

Input 2

XX mV

• Carry out the same measurements as with Input 1.

• Press .

analog input test 1...5

• Short-circuit input ‘Ind I’ (e.g. with cable 3.496.5070).

• Press <3>.

Input 1-2

XX mV

The differencial voltage between inputs ‘Ind I‘ and ‘Ind II’ is displayed.

Example: 0 - (+)1500 mV = -1500 mV

• Remove cables from the inputs ‘Ind I’ and ‘Ind II’.