6analogue input test – Metrohm 799 GPT Titrino User Manual
Page 201
![background image](/manuals/586575/201/background.png)
4.2.
Diagnosis
799 GPT Titrino
197
• Press
diagnose press key 0...9
6
Analogue input test
• Press <7>.
analog input test 1...5
6.1
Examination of highly insulated measuring inputs
• Connect ‘Ind I’ measuring input to a voltage calibrator (e.g. Metrohm 773 pH simu-
lator) by means of a highly insulated cable (e.g. 6.2108.060). Set calibrator to 0 V.
• Press <1>.
Input 1
0.0 mV
Tolerance: ± 0.5 mV
• Set the calibrator voltage, on setting 'low ohmic' (with 773 = ~ 0.002 MΩ) to differ-
ent values (e.g. +1500 mV) and compare with the displayed value.
Tolerance (with ± 1500 ÷ 2000 mV) ± 1 mV. Be aware of the calibrator's tolerance.
• Set simulator to high ohmic condition (with 773 = 1000 MΩ).
The displayed reading may vary slightly only (with 1500 mV ≤ 1 mV)
• Press
analog input test 1...5
• Disconnect simulator from ‘Ind I’ input and connect to ‘Ind II’ input.
• Press <2>.
Input 2
XX mV
• Carry out the same measurements as with Input 1.
• Press
analog input test 1...5
• Short-circuit input ‘Ind I’ (e.g. with cable 3.496.5070).
• Press <3>.
Input 1-2
XX mV
The differencial voltage between inputs ‘Ind I‘ and ‘Ind II’ is displayed.
Example: 0 - (+)1500 mV = -1500 mV
• Remove cables from the inputs ‘Ind I’ and ‘Ind II’.