2.6 Parameters, key
799 GPT Titrino
55
2.6.5 Parameters for DOS
dosing parameters
control the dosing
stop conditions:
Parameters for termination.
statistics:
Calculation of mean values and standard deviation,
see page 77.
monitoring
of limit values for the measured value and temperature.
preselections:
ON/OFF of various auxiliary functions such as auto-
matic requests after the start and activate pulse.
>dosing parameters
dispensing type: volume
volume 10.00 ml
disp.time 100 s
rate max. ml/min
disp.crit: rate
pause 0 s
**titr.
time interval 10 s
dos.element: internal D0
Dosing parameters
Type of dosing (volume, time, rate)
Main parameters.
Two parameters of the equation "rate = volume/time"
are specified, the third is calculated.
The requests which appear depend on the type of
dosing selected:
Volume (0...99 999.99 mL)
Dispensing time (1...999 999 s)
Dispensing time only. Waiting times, e.g. for refilling
are not included.
Rate (0.001...150 mL/min, max.)
Dispensing criterion (volume, time, rate)
Selection of the second dispensing parameter.
The request which appears depends on the dispensing
criterion selected.
Pause (0...999 999 s)
Waiting time, e.g. for stabilization of the electrode after
start. The waiting time can be aborted with
Time interval (1...999 999 s)
Time interval for the entry of the values in the
measuring point list. The measuring point list can
contain max. 500 points.
Selection of dosing element (internal D0, external D1,
external D2)
internal D0: internal buret of the Titrino
external D1/2: buret D1, D2, resp
parameters
>dosing parameters
>stop conditions
>statistics
>monitoring
>preselections