beautypg.com

5 parameters for dos – Metrohm 799 GPT Titrino User Manual

Page 59

background image

2.6 Parameters, key

799 GPT Titrino

55

2.6.5 Parameters for DOS









dosing parameters

control the dosing

stop conditions:

Parameters for termination.

statistics:

Calculation of mean values and standard deviation,

see page 77.

monitoring

of limit values for the measured value and temperature.

preselections:

ON/OFF of various auxiliary functions such as auto-

matic requests after the start and activate pulse.


>dosing parameters

dispensing type: volume






volume 10.00 ml

disp.time 100 s



rate max. ml/min

disp.crit: rate




pause 0 s

**titr.



time interval 10 s




dos.element: internal D0




Dosing parameters

Type of dosing (volume, time, rate)

Main parameters.

Two parameters of the equation "rate = volume/time"

are specified, the third is calculated.

The requests which appear depend on the type of

dosing selected:

Volume (0...99 999.99 mL)

Dispensing time (1...999 999 s)

Dispensing time only. Waiting times, e.g. for refilling

are not included.

Rate (0.001...150 mL/min, max.)

Dispensing criterion (volume, time, rate)

Selection of the second dispensing parameter.

The request which appears depends on the dispensing

criterion selected.

Pause (0...999 999 s)

Waiting time, e.g. for stabilization of the electrode after

start. The waiting time can be aborted with .

Time interval (1...999 999 s)

Time interval for the entry of the values in the

measuring point list. The measuring point list can

contain max. 500 points.

Selection of dosing element (internal D0, external D1,

external D2)

internal D0: internal buret of the Titrino

external D1/2: buret D1, D2, resp

parameters

>dosing parameters

>stop conditions

>statistics

>monitoring

>preselections