Metrohm 799 GPT Titrino User Manual
Page 165
3.2 Remote control commands
799 GPT Titrino
161
3.2.2.95. Config.PeriphUnit.CharSet1
Epson, Seiko, Citizen, HP, IBM
Config.PeriphUnit.CharSet2
Selection of the character set and the graphics control characters for COM1
resp. COM2 of the Titrino.
IBM means the IBM character set following character set table 437 and IBM
graphics control characters. Select 'IBM' for work with the computer.
3.2.2.96. Config.PeriphUnit.RepToComport
1, 2, 1&2
Selection of COM of the Titrino where manually triggered reports should be
outputted.
3.2.2.97. Config.PeriphUnit.Balance Sartorius,Mettler,Mettler AT,
AND,Precisa
Selection of the balance type.
3.2.2.98. Config.PeriphUnit.Stirrer
ON, OFF
Automatic stirrer control. With "ON" the stirrer will be switched on after start-
ing a method. At the end of the method it is switched off again. Valid for DET,
MET, SET, KFT, STAT, DOC.
3.2.2.99. Config.PeriphUnit.RemoteBox.Status
ON, OFF
Config.PeriphUnit.RemoteBox.Keyboard US, deutsch, francais,
español,
schweiz.
Config.PeriphUnit.RemoteBox.Barcode input, method, id1, id2,
id3, smpl size
Connections via Remote Box.
.Status:
Select if a Remote Box is connected.
.Keyboard: Type of keyboard which is connected to the Remote Box.
.Barcode: Select target in Titrino where you wish to have the string from the
barcode reader. "input" means that the string comes into the field
where the cursor is currently placed.
3.2.2.100. Config.Aux.Language english, deutsch, francais, espanol,
italiano, portugese, svenska
Selection of the dialog language.
3.2.2.101. Config.Aux.Set $G
Config.Aux.Set.Date YYYY-MM-DD
Config.Aux.Set.Time HH:MM
Date and time.
Input format of the date: Year-month-day, two-digit, enter leading zeros.
Input format for the time: Hours:minutes, two-digit, enter leading zeros.
Date and time have to be set with &Config.Aux.Set $G just after entry of the
value.