Metrohm 799 GPT Titrino User Manual
Page 15
2.4 Configuration, key
799 GPT Titrino
11
dos.element: internal D0
warn.interv.DX OFF min
dos.drive: Dosimat
Selection of the dosing element (internal D0, external
D1, external D2)
internal D0: buret of the Titrino
external D1/2: buret D1, D2, resp.
Warning interval for prep (5...9999 min, OFF)
Selection of the type of dosing element (Dosimat,
Dosino)
The internal buret D0 reacts like a "Dosimat".
volume DX 3.5 ml
cycles DX 2
dos.rate max. ml/min
fill rate max. ml/min
Parameters for the internal buret and for
Dosimats:
Volume, that will be dosed with the preparation (0...99
999.99 mL)
Number of cycles (1...9)
Dosing rate (0.01...150 mL/min, max.)
Filling rate (0.01...150 mL/min, max.)
outlet: tip
len.dos.tub.DX 40.0 cm
diam.dos.tub.DX 2.0 mm
len.asp.tub.DX 25.0 cm
diam.asp.tub.Dx 2.0 mm
dos.rate max. ml/min
fill rate max. ml/min
Parameters for Dosinos:
Location where the liquid will be expelled (tip, flask)
With "flask" the reagent will be re-expelled into the
bottle. In this case, it is possible that the titer will be
changed!
The next 4 entries serve to calculate the volumes which
will be expelled in prep:
Length of the dosing tube (1...999.9 cm)
Diameter of the dosing tube (0.1...9.9 mm)
Length of the aspirating tube (1...999.9 cm)
Diameter of the aspirating tube (0.1...9.9 mm)
Dosing rate (0.01...150 mL/min, max.)
Filling rate (0.01...150 mL/min, max.)