2.4 Configuration, key

799 GPT Titrino

11


dos.element: internal D0




warn.interv.DX OFF min

dos.drive: Dosimat

Selection of the dosing element (internal D0, external

D1, external D2)

internal D0: buret of the Titrino

external D1/2: buret D1, D2, resp.

Warning interval for prep (5...9999 min, OFF)

Selection of the type of dosing element (Dosimat,

Dosino)

The internal buret D0 reacts like a "Dosimat".




volume DX 3.5 ml


cycles DX 2

dos.rate max. ml/min

fill rate max. ml/min

Parameters for the internal buret and for

Dosimats:

Volume, that will be dosed with the preparation (0...99

999.99 mL)

Number of cycles (1...9)

Dosing rate (0.01...150 mL/min, max.)

Filling rate (0.01...150 mL/min, max.)


outlet: tip






len.dos.tub.DX 40.0 cm

diam.dos.tub.DX 2.0 mm

len.asp.tub.DX 25.0 cm

diam.asp.tub.Dx 2.0 mm

dos.rate max. ml/min

fill rate max. ml/min

Parameters for Dosinos:

Location where the liquid will be expelled (tip, flask)

With "flask" the reagent will be re-expelled into the

bottle. In this case, it is possible that the titer will be

changed!

The next 4 entries serve to calculate the volumes which

will be expelled in prep:

Length of the dosing tube (1...999.9 cm)

Diameter of the dosing tube (0.1...9.9 mm)

Length of the aspirating tube (1...999.9 cm)

Diameter of the aspirating tube (0.1...9.9 mm)

Dosing rate (0.01...150 mL/min, max.)

Filling rate (0.01...150 mL/min, max.)