Metrohm 799 GPT Titrino User Manual

Page 155

background image

3.2 Remote control commands

799 GPT Titrino

151

3.2.2.41. Mode.Parameter.TitrPara.TStart

0...999 999

Mode.Parameter.TitrPara.StartMeas pH:

0.00...±20.00, OFF

U,

Ipol:

0...±2000, OFF

Upol:

0.0...±200.0, OFF

Mode.Parameter.TitrPara.RateStart 0.01...150,

OFF

Parameters for STAT: Start of the measured point acquisition after the start

condition has been met.

.TStart: Start time in s.

.StartMeas: Start measured value.

.RateStart: Start rate in mL/min. The current rate must be less than the start

rate .This criterion does not become effective until 10 s after the

start.

3.2.2.42. Mode.Parameter.StopCond.TStop.Type abs.,rel.,delta,delay,OFF

Mode.Parameter.StopCond.TStop.Time 0...999 999

Mode.Parameter.StopCond.TStop.Factor 0...±999 99

Mode.Parameter.StopCond.TStop.Delta 0...999 999

Mode.Parameter.StopCond.TStop.Delay 0...999 999

Parameters for STAT: Stop time.

If an absolute stop time (abs.) has been selected, the time is in s (Time).

A relative stop time (rel.) is calculated as a function of the sample size:

Stop time in s = sample size * factor

The factor selected (Factor) applies.

delta: Time in s after control point is reached for the first time.

delay: Time in s after last dosing step.

OFF means that the criterion is not monitored.

3.2.2.43. Mode.Parameter.StopCond.RateStop 0.01...150, OFF

Parameters for STAT: Stop rate in ml/min. The current rate must be less than

the start rate. This criterion does not become effective until 10 s after the

start.

3.2.2.44. Mode.Parameter.Evaluation.TimeWin.1.LowLim 0...999 999, OFF

Mode.Parameter.Evaluation.TimeWin.1.UpLim 0...999

999,

OFF

Parameters for STAT: Evaluation of rates in time windows. Entry of the times

in s. The evaluated rates are available as variables C8X (X = 1...9).

3.2.2.45. Mode.Parameter.Evaluation.FixVol.1.Value 0...999

999,

OFF

Parameters for STAT: Interpolation of the volume at preset times. Entry of the

times in s. The interpolated volumes are available as variables C5X.

3.2.2.46. Mode.Parameter.Evaluation.FixTime.1.Value 0.01...1.00,

OFF

Parameters for STAT: Interpolation of the time at preset fraction of the final

volume. Entry as part of V(tot). The interpolated volumes are available as vari-

able C6X.