background image

4.1. Error and special messages

799 GPT Titrino

192

Error messages in connection with the data transfer

If neither a computer nor a printer is attached, the report output at the end of the titration

must be switched off:

Receive errors:

error 36

Parity

Exit: and set corresponding quantity the

same on both instruments

error 37

Stop

bit

Exit: and set corresponding quantity the

same on both instruments

error 38

Overrun error. At least 1 character could not be read.

Exit:

error 39

Overflow of the receive buffer of the Titrino (> 128

characters).

Exit:

Send errors:

error 42

CTS=OFF

Handshake unsatisfactory for more than 1 s.

Exit: Is the receiver switched on and ready

to receive?

error 43

The transmission of the Titrino has been interrupted

with XOFF for at least 6 s.

Exit: .

error 45

The receive buffer of the Titrino contains an

incomplete string (missing L

F

). Transmission of the

Titrino is thus blocked.

Exit: Send

L

F

or .