2.4 Configuration, key

799 GPT Titrino

13

>curve





grid: ON


frame: ON


scaling: Full





width 0.75



length 0.1

curve

The settings are valid for COM1 and COM2.

If you change the printer type, the following settings are

initialized according to the printer.

Grid drawing (ON, OFF)

Frame drawing (ON, OFF)

Type of scaling (Full, Auto)

Full: The scaling goes from the greatest to the smallest

value.

auto: The scaling from tick to tick, e.g. the

smallest/greatest values lie in between the first/last tick.

Width (0.2...1.00)

1 is greatest width. If you set 1 you may loose the label

at the right margin.

Length (0.01...1.00)

Volume axis

Curve length per cylinder volume

0.01 100

cm

0.1 10

cm

0.5 2

cm

1 1

cm

Time axis:

Curve

length

0.01 100

cm

0.1 10

cm

0.5

2 cm

1 1

cm