2.7 Result calculations
799 GPT Titrino
75
RS1 text RS1
RS1 decimal places 2
RS2 unit: %
RS1 limit control: OFF
RS1 low lim. 0.0
RS1 up lim. 0.0
RS1 L13 output: OFF
Text for result output (up to 8 characters)
Text input see page 6.
Number of decimal places for result (0...5)
Selection of result unit (%, ppm, g/L, mg/mL, mol/L,
mmol/L, g, mg, mL, mg/pc, s, mL/min, no unit or up to
6 characters).
Limit control for the result (on, off)
The limits are checked each time a result is calculated.
If "on" has been set:
Lower limit (0.0...999 999)
Upper limit (0.0...999 999)
Sets line L13 of the remote socket (OFF, active, pulse)
if the result lies outside the limits.
Enter next formula, e.g. for RS2.
Meaning of the calculation variables CXX:
C00
Sample size, see page 95.
C01...C19
Method specific operands, see page 76. They are stored with the method
in the method memory.
C21...C23
Sample specific operands, see page 95ff.
C26, 27
Mean values from silo calculations.
C30...C39 Common
variables.
C40
Initial measured value of the sample, last measured value in MEAS.
C41 End
volume.
C42 Determination
time.
C43
Volume drift for KFT and SET with conditioning.
C44 Temperature.
C45
Dispensed start volume.
C46 Asymmetry-pH
(calibration).
C47
Electrode slope (calibration).
C48
Volume value at maximum voltage in the curve (for STAT, DOS, DOC
measured value monitoring has to be active; no evaluation in CAL and
TIP).
C49
Volume value at minimum voltage in the curve (for STAT, DOS, DOC
measured value monitoring has to be active; no evaluation in CAL and
TIP).
C51...C59
Fix EP for DET and MET or Fix V for STAT.
C61...C69
pK/HNP values for DET and MET or Fix V for STAT.
C70...C79
Temporary variables for calculations in TIP.
C80
Mean rate for STAT.
C81...C89
Rates, evaluated in preset time windows for STAT.