1 manual dosing, 2 preparation of the titration burets, key

Page 107

background image

2.18 Manual dosing and preparation of titration burets

799 GPT Titrino

103

2.18 Manual dosing and preparation of titration burets

2.18.1 Manual dosing

serves as preselection for the buret: internal

D0, external D1, external D2.

With , the preselected buret will dose as long

as is pressed. The dosing rate can be set with

the analog potentiometer at the Titrino. If no buret has

been preselected, the buret which is active in the

method, will dose.

2.18.2 Preparation of the titration burets, key

The titrating burets should be prepared, if
• they have not been used for a long time

• a new bottle of reagent is used

• you are using the exchange unit for the first time.

internal D0 prep

external D1: prep

With , the preparation (or emptying, resp.) of

the selected buret will begin. The parameters for the

preparation are under , see page 10.

To change the buret, press repeatedly.

Attention: Liquid will be expelled at the buret tip!

Internal buret D0, Titrino buret

External buret D1

The Titrino recognizes the buret type automatically

(Dosimat or Dosino).

For Dosinos:
• After the preparation the dosing unit is without air

bubbles and filled up to the top of tube.

• Press <←/→> for "empty" . This function empties

the Dosino completely, e.g. for cleaning the dosing

unit.

External buret D2

as for buret D1

Prep sequence for Dosinos:

The following volumes are dosed:

Volume of the aspiration tube, volume of the cylinder, volume of the dosing tube.

The cylinder volume can be expelled either through the top or back to the bottle. The

parameters are entered under , see page 11.

PREP

RS

4

DOS