Metrohm 799 GPT Titrino User Manual
Page 118

3.1. General rules
799 GPT Titrino
114
E171
Rate too low in DOS. No dispensing possible with the Exchange Unit currently
mounted.
Exit: The error message disappears on next start or &m $S.
E172
In TIP or DOS a QuickMeas was started, without defining a measuring
quantity.
Exit: The error message disappears on next start or &Mode.QuickMeas $S.
E173
The warning interval of the internal buret D0 called.
Exit: Execute prep &a.b.p $G or start next method.
E174
The warning interval of the external buret D1 called.
Exit: Execute prep &a.b.p $G or start next method.
E175
The warning interval of the external buret D2 called.
Exit: Execute prep &a.b.p $G or start next method.
E176
The function &Assembly.Buret.Prep or &Assembly.Buret.Empty was
interrupted manually.
Exit. The error message disappears on next start.
E177
Accessing to the memory card, the card was not (properly) inserted.
Exit. The error message disappears on next start.
E178
The date of changing the battery of the card is expired.
Exit. The error message disappears on next start.
E180
Memory card write-protected.
Exit: Send new command.
E181
Memory card not formatted.
Exit: Send new command.
E182
Memory card not accessible.
Exit: Send new command.
E183
A directory with the same name exists already on the memory card.
Exit: Send new command.
E184
Measured value below lower limit.
Exit: The error message disappears when again within the limits or &m $S.
E185
Measured value above upper limit.
Exit: The error message disappears when again within the limits or &m $S.
E186
Temperature below lower limit.
Exit: The error message disappears when again within the limits or &m $S.
E187
Temperature above upper limit.
Exit: The error message disappears when again within the limits or &m $S.
E188
Rate below lower limit.
Exit: The error message disappears when again within the limits or &m $S.
E189
Rate above upper limit.
Exit: The error message disappears when again within the limits or &m $S.
E196
Result is out of limits.
Exit: The error message disappears on next start or on recalculation.
E197
Sample size is out of limits.
Exit: The error message disappears on next start or on introduction of new
sample size.
E198
Validation interval is expired.
Exit: The error message disappears on next start or clear counter with
&Config.Monitoring.Validation.ClearCount $G.
E199
Service date is reached.
Exit: The error message disappears on next start or change date in
&Config.Monitoring.Service.Date.