Model 493.40 i/o carrier connections, Model 493.40 i/o carrier connections 87 – MTS Model FlexTest SE User Manual
Page 87
![background image](/manuals/568034/87/background.png)
Model 493.40 I/O Carrier Connections
Models FlexTest® IIm/GT/SE Controller Hardware
FlexTest GT Controller Connections
87
Model 493.40 I/O Carrier Connections
Each Model 493.40 I/O Carrier module can include up to four mezzanine cards.
Each installed mezzanine card is assigned a specific I/O Carrier module front
panel connector (J4–J7 I/O).
A hardware interface file (hwi) defines each type of module (and their associated
mezzanine cards) and maps each module location for the system software. The
.hwi file and the physical locations for each type of module and associated
mezzanine cards must match. Also, the I/O Carrier module address setting (on
the module) must match the .hwi file address. For more information on the .hwi
file, see the MTS 793 Utility Software manual.
•
Transducer connections require a conditioner mezzanine card to be installed
in the I/O Carrier module. The following conditioners can be installed:
–
Model 493.21B Digital Universal Conditioner
–
Model 493.25 Digital Universal Conditioner
–
Model 493.47 Encoder
–
Model 493.48 Acceleration Conditioner
•
Valve connections require a valve driver mezzanine card be installed in the
I/O Carrier module. The following valve drivers can be installed:
–
Model 493.14 Valve Driver
–
Model 493.15 3-Stage Valve Driver
•
Analog I/O connections require an A/D or D/A mezzanine card be installed
in the I/O Carrier module. The following analog mezzanine cards can be
installed:
–
Model 493.45 A/D Converter
–
Model 493.46 D/A Converter