MTS Model FlexTest SE User Manual
Page 68
Models FlexTest® IIm/GT/SE Controller Hardware
Transition Module Connectors
Installation
68
Single chassis
configuration
If you have a single analog chassis:
First install a cable from a Series 498 host computer serial port (debug) to J51
Intlk In on the Model 497.05 Hydraulic Control Panel, then from J52 Intlk Out
to J51 Intlk In on the Model 497.01 Analog Chassis. Then install jumper plug
assembly (PN 046-350-701) to J52 Intlk Out on the analog chassis.
Multichassis
configuration
If you have a multiple analog chassis:
1. Install the cable from the chassis to J51 Intlk In on the Model 497.05
Hydraulic Control Panel, then from J52 Intlk Out to J51 Intlk In on the
first Model 497.01 Analog Chassis (see
2. Install a chassis interconnect cable from J52 Intlk Out of the first analog
chassis to J51 Intlk In of the next chassis. Install as many cables as needed
to connect all the chassis.
3. On the last chassis, install jumper plug assembly (PN 046-350-701) to J52
Intlk Out.