MTS Model FlexTest SE User Manual
Page 161
Connecting Interlock Signals to 407 Controllers
Models FlexTest® IIm/GT/SE Controller Hardware
FlexTest GT Controller Connections
161
How to connect
hydraulic terminals on
master-dependent 407
Controllers
The following figure shows how to cable hydraulic terminals on three Model 407
Controllers in a master-dependent daisy-chain. Note that the last dependent
controller in any daisy-chain must have jumper plugs installed in its Hyd Out
and Intlk In terminals.
Jumper Plug PN
049-635-901
Jumper Plug PN
049-635-901
Hyd
In
Hyd
Out
Intlk
In
Intlk
Out
Dependant 407 Controller
Hyd
In
Hyd
Out
Intlk
In
Intlk
Out
Last Dependant
407 Controller
Hyd
In
Hyd
Out
Intlk
In
Intlk
Out
Master 407 Controller
No Connection
No Connection