152
Do not interrupt Model 493.10 Chassis power when you are programming
another controller. A power interrupt may cause the external controller to
react to a zero command and move the actuator unexpectedly.
Ensure that power cannot be removed accidently while a test is in progress.
For more detailed information about analog I/O connections, see
WARNING
407 Controller
Ch 1
Ch 2
Ch 3
Ch 4
Ch 5
Ch 6
m
J11
(Ch 1-4)
J12
(Ch 5-6)
493.77
Filtered In
PN 044-632-0XX
Ch 1
Ch 2
Ch 3
Ch 4
Ch 5
Ch 6
m
J11
(Ch 1-4)
J12
(Ch 5-6)
493.76
Analog Out
493.76 Analog Out
This module includes six
BNC connectors that
provide analog outputs from
a Model 493.46 D/A
mezzanine card.
Six analog outputs are
available with each D/A
mezzanine card installed on
the Model 493.40 I/O
Carrier module.
493.77 Filtered
Analog Input
This module includes
six BNC connectors
that provide filtered
analog inputs to a
Model 493.45 A/D
mezzanine card.
Six analog inputs are
available with each A/D
mezzanine card
installed on the Model
493.40 I/O Carrier
module.
Transition modules on the
rear panel of the Model
493.10 chassis are used for
analog inputs and outputs.
Single 407 Controller Connections