Connecting Electrical Power
Models FlexTest® IIm/GT/SE Controller Hardware
Installation
53
Connecting Electrical Power
Electrical connections must be made by qualified personnel and conform to local
codes and regulations. An electrical service panel to provide the electrical power
feed (line voltage) to the chassis is not necessary, but may be required by local
electrical codes.
Note
Local electrical codes supersede any information found here.
Grounding
The chassis will not function correctly if it is not grounded as shown. Be sure
your power source is also properly grounded. The chassis includes two grounds:
a chassis ground and a signal ground. The two grounding lugs are connected
together with an external shorting bar when the chassis is manufactured.
•
For the console configuration, remove the shorting bar from the ground lugs
and connect the chassis ground to the console rail.
•
For a stand-alone configuration, always connect the shorting bar to both
ground lugs.
Chassis
Ground
Signal
Ground
System ground cable
(PN 037-766-102)
connected to other
components.
Grounding the
chassis
TYPE E PC-MIP
T YP E E PC-MIP
RUN 6TS BF L
ABT
RST
DEBUG
PCI MEZZANNE CARD
Power
100–240 VAC
50–60 Hz, 12–6A
TS-G435
Chassis ground is
connected through the
power cord connector.