Independent hsm (no hpu), Independent hsm (no hpu) 246 – MTS Model FlexTest SE User Manual
Page 246
![background image](/manuals/568034/246/background.png)
Models FlexTest® IIm/GT/SE Controller Hardware
Model 493.02 Chassis Multiple Controller
Hydraulic Configurations
246
Model 493.02 Chassis Multiple Controller Interconnections
FlexTest SE Controllers support the following hydraulic interconnections for
multiple controller configurations:
•
Independent HSM control with no HPU
•
Independent HSM control with a shared HPU
•
Shared HSM with an HPU
•
Shared HSM without HPU
•
Independent HSM controlled with a shared HPU using first-on-last-off
For more detailed information about multiple controller configurations,
see
“Multiple Controller Connections”
on page 220 and the FlexTest SE User
Manual.
Independent HSM (no HPU)
For this multiple controller configuration:
•
Each controller has an independent hydraulic service manifold (HSM).
•
No hydraulic power unit (HPU) is connected.
•
Select HSM Only for Power Type on each connected controller.
Path: Config > Edit Hardware > Power Options > Power Type > HSM
Only
FlexTest SE
Emergency Stop
7
8
9
6
5
4
3
2
1
.
0
+/-
enter
cancel
FG
Setup
Status
Scope
Meters
Tuning
Config
Limits
Log
Menu
?
A
B
Recall
Enabled
reset
Rewind
Run
Hold
Stop
Power
Program
Hydraulic
Interlock
Program
Interlock
HPU
HSM
Off
Low
High
Off
Low
High
Navigate
Monitor 1
Monitor 2
FlexTest SE
Emergency Stop
7
8
9
6
5
4
3
2
1
.
0
+/-
enter
cancel
FG
Setup
Status
Scope
Meters
Tuning
Config
Limits
Log
Menu
?
A
B
Recall
Enabled
reset
Rewind
Run
Hold
Stop
Power
Program
Hydraulic
Interlock
Program
Interlock
HPU
HSM
Off
Low
High
Off
Low
High
Navigate
Monitor 1
Monitor 2
Power Type = HSM Only
Power Type = HSM Only
Interlock/Hydraulics
HSM 1
056-126-4xx
J28 HSM
HSM 2
056-126-4xx
J28 HSM
056-316-0xx