J29 load unit, J29 load unit 136, Cable specification – MTS Model FlexTest SE User Manual
Page 136: Jumper plug, Station stop
![background image](/manuals/568034/136/background.png)
Models FlexTest® IIm/GT/SE Controller Hardware
J29 Load Unit
FlexTest GT Controller Connections
136
J29 Load Unit
Connector J29 Load Unit connects to the load unit lift/lock panel.
Cable specification
•
P29 is a 15-contact, type D, male, EMI connector.
•
Cable for load frames with crosshead locks built after 1985, 18 AWG, 8
conductor with overall foil shield Alpha 5388C or equivalent with drain
wire connected to pin 9 at both ends of the cable.
•
Cable for all load frames without crosshead locks–22 AWG, 6 conductor
with overall foil shield Alpha 5386C or equivalent with drain wire
connected to pin 9 at both ends of the cable (pin 9 may be pin E at the load
frame).
Jumper plug
If connector J29 is not used, you must install a jumper plug to maintain the
integrity of the interlocks. Use jumper plug 100-007-947 or jumper pins: 3 and 4;
5 and 7; 8 and 13.
Station stop
The Emergency Stop connection can also be configured as a station stop. When
this is done, be sure that you have other Emergency Stop boxes near by. Pressing
Station Stop will shut down the hydraulics to an individual station without
shutting down power to the hydraulic power unit.
Note
A Model 493.73 HPU Transition Board must be installed for the
Emergency Stop connection to be enabled.
+24 V DC
+24 V DC
OR
Station
Stop
Emergency
Stop
Cross
Head
Lock
1
4
3
2
13
8
7
5
9
12
14
Cross Head Unlock
DC Common
HSM hi
Crosshead
Solenoid
Shield
+24 V
E-Stop
CRM Relay
in HPS
X
*
X
Pins 1, 2, 3, and 4 can be
configured for contacts (with
jumpers X
*
and X
†
) or a logic
signal (without jumpers X
*
and
X
†
).
X
*
= Jumper X2 for Channel A
Jumper X1 for Channel B
X
†
= Jumper X4 for Channel A
Jumper X3 for Channel B
Pins 5, 7, 8, and 13 maintain
the continuity of the
emergency stop interlock.