Adda ii connections, Adda ii connections 124 – MTS Model FlexTest SE User Manual
Page 124
![background image](/manuals/568034/124/background.png)
Models FlexTest® IIm/GT/SE Controller Hardware
ADDA II Connections
FlexTest GT Controller Connections
124
ADDA II Connections
Each optional Model 493.50 ADDA II module supports up to four mezzanine
cards. Each installed mezzanine card is assigned a specific ADDA II module
front panel connector (J11–J18).
A hardware interface file (.hwi) defines each type of module (and their associated
mezzanine cards) and maps each module location for the system software. The
hwi file and the physical locations for each type of module and associated
mezzanine cards must match. Also, the ADDA II module address setting (on the
module) must match the .hwi file address. For more information on the .hwi file,
see the MTS 793 Utility Software manual.
The following A/D or D/A mezzanine cards can be installed:
–
Model 493.55 A/D
–
Model 493.56 D/A
–
Model 493.57 DSPAD
–
Model 493.59 Universal Encoder