J44 run/stop, J44 run/stop 138, Run/stop status – MTS Model FlexTest SE User Manual
Page 138: Interlock status
![background image](/manuals/568034/138/background.png)
Models FlexTest® IIm/GT/SE Controller Hardware
J44 Run/Stop
FlexTest GT Controller Connections
138
J44 Run/Stop
Run/Stop status
Connector J44 Run/Stop provides the run/stop status of the controller to external
devices.
•
Two form C contacts provide the run/stop status.
•
The contacts are rated 1.0 A at 30 V (AC or DC).
Interlock status
Connector J44 Run/Stop also provides opto-isolator outputs that indicate the
interlock status of each station to an external device. These interlock status
outputs are normally on and will turn off when an interlock occurs. For detailed
circuit drawings, see
Run/Stop 1
+
-
Interlock Status
Output
1
2
3
4
5
6
7
8
Run/Stop 0
J44
To
External Devices
From
Chassis