J52 box out, J52 box out 175, Last controller jumper – MTS Model FlexTest SE User Manual
Page 175: Cable specification
J52 Box Out
Models FlexTest® IIm/GT/SE Controller Hardware
FlexTest GT Controller Connections
175
J52 Box Out
The J52 Box Out connector provides an interface for connecting multiple MTS
controller boxes together to allow sharing of station interlock status and a master
synchronization clock.
Last controller jumper
The last controller box in a chain must have a jumper plug (MTS PN 100-079-
126) on its J52 Box Out connector. This jumper allows the return of the
composite station interlock signal for all controllers in the chain to monitor.
Note
If there is only one controller, a jumper is still required on its Box Out
connector.
Cable specification
Use MTS cable PN 056-534-1xx.
+clk in
/intlk out 2
/intlk out 1
/intlk out 3
/intlk out 4
test
-clk out
/intlk return in 1
/intlk return in 2
/intlk return in 3
/intlk return in 4
signal return
signal return
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
J52 Box Out