Interlock Jumper Plugs
Models FlexTest® IIm/GT/SE Controller Hardware
FlexTest IIm Controller Connections
193
Interlock Jumper Plugs
Several jumper plugs must be installed to maintain the integrity of the interlock
chain. Each jumper plug represents a place where you can add a custom
connection to the interlock chain.
Model 497.05
Hydraulic Control
Panel
Any interlocks connected to the hydraulic control panel will generate a hydraulic
interlock which shuts down the hydraulic power supply. This type of interlock
shuts down all test stations.
•
Connector J52 requires a jumper plug to close the interlock chain.
•
If J26 is not connected, it needs a jumper plug.
•
Connectors J15A–J17C (Remote Interlock Inputs) each require a jumper
plug unless it is being used. Jumper both pins of each connector.
Series 498 System
Digital I/O
Any interlock connected to the Series 498 System Digital I/O transition module
can generate a station interlock to disable pressure (the HSM) to that station.
The Series 498 System Digital I/O transition module supports:
•
Four stations of Hydraulics On digital inputs
•
Four stations of Hydraulics Hi digital inputs
•
Four stations of Hydraulics Enables
•
Four stations of Remote Stop/Run outputs
•
Four stations of program interlocks
•
Four stations of auxiliary interlocks
•
Four stations of 497 interlock inputs
•
Four stations of 497 interlock outputs
•
Four stations of auxiliary interlock inputs
•
Four stations of auxiliary interlock outputs
•
Two spare digital inputs
•
Two spare digital outputs