J29 Emergency Stop Connections
FlexTest SE Controller Connections
232
J29 Emergency Stop Connections
Connector J29 E-STOP provides an output to external devices when an
emergency stop signal is generated. This connector may also be used with an
external E-stop.
Cable specification
•
15 contact type D male EMI connector.
•
Backshell–EMI metallized plastic.
•
Cable–24 AWG 4 connector with braided shield, with the braid connected to
a metallized plastic backshell at the chassis and to ground at the emergency
stop station.
J29 Connector Pin Assignments
P
IN
N
AME
F
UNCTION
1
xhead 24V
Fused +24V
2
xhead intlk com
24V common
3
Program interlock low
(xhead interlock low)
Low side of opto-isolator, jumpered to common
4
Program interlock high
(xhead interlock high)
High side of opto-coupler, pulled up to 24V (opto on = intlk)
5
Local E stop
Local E stop loop (connected to pin 7)
6
N/C
No connection
7
Local E stop
Local E stop loop (connected to pin 5)
8
HPU E stop
E stop loop from HPU (connect to pin 13)
9
N/C
No connection
10
N/C
No connection
11
N/C
No connection
12
xhead power
+24 V crosshead unlock power output (from HSM high relay
contact)
13
HPU E stop
E stop loop from HPU (connect to pin 8)
14
xhead power return
24 V common for crosshead unlock power output
15
N/C