Analog inputs – MTS Model FlexTest SE User Manual
Page 186

Models FlexTest® IIm/GT/SE Controller Hardware
ADDA II Connections
FlexTest IIm Controller Connections
186
Analog inputs
The analog-to-digital mezzanine cards accommodate up to eight analog input
signals. Each A/D input signal must be within ±10 volts.
•
Analog inputs can be connected at the BNC connectors of a Model 493.81
Analog In transition module and the outputs from this module are connected
to the Model 493.50 ADDA II module. This configuration is typically used
for DSPAD mezzanine cards that provide on-board digital filtering.
•
Analog inputs can be connected at the BNC connectors of a Model 493.83
Filtered Analog Input transition module and the outputs from this module
are connected to the Model 493.50 ADDA II module. This configuration is
typically used for A/D mezzanine cards that requires filtering.
Input 1
Input 2
Input 3
Input 4
J 1 1 , J 1 3
J 1 5 , J 1 7
J 1 2 , J 1 4
J 1 6 , J 1 8
+
-
+
-
+
-
+
-
+
-
+
-
+
-
+
-
Input 5
Input 6
Input 7
Input 8
To
A/D
From
493.81 Analog In
Module
1
2
3
4
5
6
7
8
1
2
3
4
5
6
7
8