Chassis maintenance, Appendix c, Chassis maintenance 253 – MTS Model FlexTest SE User Manual
Page 253
Models FlexTest® IIm/GT/SE Controller Hardware
Chassis Maintenance
253
Appendix C
Chassis Maintenance
This section describes how to maintain your MTS controller chassis.
Cleaning the chassis
Remove any dust from the chassis with Endust for Electronics or equivalent.
Cleaning the air filter
Be sure the cooling fan is operational and not clogged. Clean or replace the filter
as required.
The filter is typically located in the top of the chassis, and can be accessed from
the rear.
To clean the air filter:
1. Remove the filter from the chassis.
2. Immerse the filter in a large pan filled with warm water and a mild
detergent, and allow it to soak a few minutes.
3. Gently agitate the filter to loosen debris.
4. Rinse the filter thoroughly with clean water.
Note
Do not use compressed air to help dry the filter. Compressed air may
damage the filter element.
5. Allow to drip until completely dry.
Note
The air filter has a distinct top and bottom. Ensure that you orient the
filter properly.
6. Replace the filter in the chassis.