Remote Station Controller (RSC) Connections
Models FlexTest® IIm/GT/SE Controller Hardware
FlexTest IIm Controller Connections
203
Remote Station Controller (RSC) Connections
When equipped with a GRES III plug-in module and a corresponding Series
498 RS-485 transition module, your FlexTest IIm controller can
accommodate up to four remote station controllers (one RSC for each
station).
Each RSC connects to a J50 serial connector on the Series 498 RS-485
transition module. The hwi file defines which connector is used for each
station.
.hwi file additions
In order to use one or more RSCs with your FlexTest controller, two sections
must be added to your .hwi file—an RSC section and a GRES III section.
Note
If you specified RSC support when you ordered your test system,
these sections will be added to your hwi file at the factory.
Note
For information on editing your .hwi file to accommodate RSCs, see
the Hwi File Editor manual.
m
498
RS-485
TRANSITION
MODULE
+12V +5V
J50A
J50B
J50C
J50D
J7 SERIAL
498 RS-485 Transition Module
Cable the RSC to the proper J50 serial
connector (usually J50A) using cable PN
055-634-3XX.
Off
Low
Interlock
HSM
Test
Manual Control
Emergency Stop
F1
F2
F3
F4
Status
Remote Station Control
Enable /
Disable*
Reset
High
Hold
Stop
Run
Off
Low
High
HPS
055-634-3XX
Remote Station Controller
Use Jumper Plug 055-634-101