D/a connections, Adda ii universal encoder connections – MTS Model FlexTest SE User Manual
Page 187
![background image](/manuals/568034/187/background.png)
ADDA II Connections
Models FlexTest® IIm/GT/SE Controller Hardware
FlexTest IIm Controller Connections
187
D/A connections
Digital-to-analog mezzanine cards support up to eight program or readout signals
to external devices. Each D/A signal is an analog output within ±10 volts.
•
Each readout signal is from a 16 bit digital to analog converter.
•
Each output is a ±10 V analog output.
•
The Model 493.56 D/A mezzanine card provides analog signals to the
Model 493.82 Analog Out transition module. Analog outputs are available
at the BNC connectors of a Model 493.82 Analog Out transition module for
external devices (DVM, oscilloscope).
ADDA II Universal
Encoder connections
Each encoder requires a Model 493.59 Universal Encoder mezzanine card
installed on the Model 493.50 ADDA II module.
Output 8
Output 7
Output 6
Output 5
-
+
-
+
-
+
-
+
Output 4
Output 3
Output 2
Output 1
-
+
-
+
-
+
-
+
J 1 1 , J 1 3
J 1 5 , J 1 7
J 1 2 , J 1 4
J 1 6 , J 1 8
From
D/A
To
External
Device
1
2
3
4
5
6
7
8
9
1 0
1 1
1 2
1 3
1 4
1 5
1 6