Station configuration, Station configuration 256 – MTS Model FlexTest SE User Manual
Page 256

Models FlexTest® IIm/GT/SE Controller Hardware
8-Station Configuration
Optional Station Configurations
256
8-Station Configuration
The 8-Station configuration can provide eight channels of control.
A typical 8-Channel/8-Station configuration requires 16 Digital Universal
Conditioners and 8 two-stage valve drivers.
HSM power limits
HSM power current is limited to 1.5 A per HSM.
Interlocks
For 8-station configurations the hwi file must contain the line
INTERLOCKS=8. This line must be a discrete entry, not part of any other hwi
section.
Cross-head interlocks with solenoid power are not supported for 8-station
configurations.
RSC
Remote Station Control (RSC) is not supported for 8-station configurations.
A/D analog inputs
Only one set of six A/D analog inputs is supported for 8-station configurations.