MTS Model FlexTest SE User Manual
Page 201

Encoder/Temposonics Connections
Models FlexTest® IIm/GT/SE Controller Hardware
FlexTest IIm Controller Connections
201
Jumper configurations
The Series 498 transition module can be configured for a Temposonics sensor, a
5 V DC encoder, a 12 V DC encoder, a 24 V DC encoder. The following is a list
of the configuration drawings:
•
049-035-301 Absolute Encoder
•
049-035-401 +5 Volt Encoder
•
049-035-501 +12 Volt Encoder
•
049-035-601 +24 Volt Encoder
•
049-035-701 Temposonics II
•
053-070-201 Temposonics III
ADDA mezzanine
cards
The ADDA module requires special mezzanine cards to support encoders and
Temposonics sensors.
Mezzanine cards for encoders and Temposonics sensors support four input
channels.
Mezzanine cards for multiturn encoders (Heidenhain 417/425) and Temposonics
III position sensors support two channels.)
•
The encoder mezzanine card counts pulses from the encoder to determine
the displacement of an actuator.
•
The Temposonics mezzanine card interprets the pulses from the
Temposonics sensor to determine the displacement of an actuator.
Optional ADDA II
configuration
For the optional ADDA II configuration (using the Model 493.50 ADDA II
module), each encoder requires installation of a Model 493.59 Universal Encoder
mezzanine card on the Model 493.50 ADDA II module.