INFICON XTC/C Thin Film Deposition Controller User Manual
Page 81
3 - 17
IP
N 07
4-
18
3X
XTC/C - XTC/2 Operating Manual
12
11,12
END Deposit
When FINAL THK is
exceeded for two
consecutive
measurements
Entry of READY
state
** Function may be overwritten by Remote Communications Commands R15 - R18,
see
.
Table 3-5 Open Collector Outputs* (one of eight encoding)
Output #
Low
High
1
18
Crucible Select 1
If the active layer's designated
crucible is 1, or 0
Balance
2
19
Crucible Select 2
If the active layer's designated
crucible is 2
Balance
3
20
Crucible Select 3
If the active layer's designated
crucible is 3
Balance
4
21
Crucible Select 4
If the active layer's designated
crucible is 4
Balance
5
22
Crucible Select 5
If the active layer's designated
crucible is 5
Balance
6
23
Crucible Select 6
If the active layer's designated
crucible is 6
Balance
7
24
Crucible Select 7
If the active layer's designated
crucible is 7
Balance
8
25
Crucible Select 8
If the active layer's designated
crucible is 8
Balance
* The crucible select outputs are open collector type, 5 volt maximum with a capability
of sinking 5 TTL loads (10 mA)
Table 3-4 Aux I/O Connector (continued)
Relay #
Pin #
Function**
Closed Contacts
Open Contacts