INFICON XTC/C Thin Film Deposition Controller User Manual
Page 112

4 - 8
IP
N 07
4-
18
3X
XTC/C - XTC/2 Operating Manual
The following variable parameters and their limits are listed below. If a value
outside the stated limits is attempted, the message ERR 1 is displayed.
Notes:
1. If configured for dual sensor, TOOL FACT 1 applies to primary sensor,
TOOL FACT 2 applies to secondary sensor. If configured for single or CrystalSix
sensor(s), TOOL FACT 1 is used regardles of whether sensor 1 or 2 is active.
2. Set to 0 for single crucible operation. If non-zero, instrument waits until Crucible Valid
input goes low before going into Rise.
Table 4-2 Limits for Film Parameters
PARAMETER
LIMITS
UNITS
Rise Time 1, 2
00:00 - 99:59
MIN:SEC
Soak PWR 1, 2
0.0 - 100
%
Soak Time 1, 2
00:00 - 99:59
MIN:SEC
Shutr DLY
Yes or No
--
NEW RATE
0.000 - 999.9
KÅ
R RAMP TIME
00:00 - 99:59
MIN:SEC
IDLE RAMP
00:00 - 99:59
MIN:SEC
IDLE PWR
0.0 - 100
%
TIME PWR
Yes or No
--
XTAL SWCH S, Q
0 - 9
Whole Numbers
TOOL FACT 1,2
10.0 - 500
% (see note 1)
DEP RATE
0.000 - 999.9
Å/SEC
FINAL THK
0.000 - 999.9
KÅ
THK SPT
0.000 - 999.9
KÅ
DENSITY
0.500 - 99.99
GM/CC
Z-RATIO
0.100 - 9.999
--
SENSOR #
1 or 2
--
SOURCE #
1 or 2
--
CRUCIBLE #
0 - 8
Whole Numbers (see note 2)
CTL GAIN
00.01 - 100.0
(Å/SEC)/%
CTL TC
0.1 - 100
SEC
CTL DT
0.1 - 100
SEC
MAXPWR
0.0 - 100
%
SAMPLE
0 - 99
%
HOLD TIME
00:00 - 99:59
MIN:SEC