8.4 Crystal Sensor Emulator
IPN 760-601-G1 or 760-601-G2
NOTE: 760-601-G2 is fully compatible with all Thin Film Deposition Controllers.
Crystal Sensor Emulator 760-601-G1 (obsolete) is not compatible for use
with an IC/5, IC/4 or Cygnus.
The Crystal Sensor Emulator option is used in conjunction with the Thin Film
Deposition Controller to rapidly diagnose problems with the Deposition Controller's
measurement system. See
.
Figure 8-5 Crystal Sensor Emulator
The Crystal Sensor Emulator may be attached at various points in the
measurement system, from the oscillator to the sensor head. It provides a known
"good" monitor crystal with known "good" electrical connections. Using the
emulator and the controller in a systematic manner provides a fast means of
isolating measurement system, cable, or sensor problems. See
.
Figure 8-6 Crystal Sensor Emulator Attachment Points