5.4.4.2 Status General Commands

The message format is:

SG|

Command ID =

Action Value = See

Table 5-9

.

Table 5-9 Status General Commands

SG

Command

ID

Meaning

Action

Value

Response

0

Data at Timer
Value

=
Requested
Timer Value

108 Bytes. Six sets of channel data at the

requested timer tick:

Sensor Status (Byte)
State (Byte)
Rate (filtered) Float
Thickness (Float)
Power (Float)
Rate Deviation (Float)
If data is not available for the timer tick

requested, then the return will be "No
Data" error. (1 byte)

1

(0x01)

Which channels
running

1 Byte, Each bit represents a channel, a bit
set means channel running (i.e. not in Ready
or Stop). MSB is channel 1. Two LSB unused.

2

(0x02)

Output Status

5 Bytes. Each bit represents an output, bit set
means output active. MSB of first byte is
output 1. Two LSB of last byte unused.

3

(0x03)

Input Status

4 Bytes. Each bit represents an input. bit set
means input active. MSB of first byte is input
1. Four LSB of last byte unused.

4

(0x04)

Logic statement
status

13 Bytes. Each bit represents a logic
statement, bit set means logic statement is
true. MSB of first byte is logic statement 1.
Four LSB of last byte unused.