INFICON Cygnus Thin Film Deposition Controller User Manual
Page 170

8 - 4
IP
N 07
4-
37
9-
P1
K
Cygnus Operating Manual
MEM ALLOC FAIL
This message appears if a Data Log or Print Screen operation fails because of
an overuse of system memory. If this message appears contact the INFICON
Service Department.
MIN POWER
Indicates the specified minimum power has been reached on an active
Channel. The Min Power message will only be displayed when the Channel is
at the minimum power level and the Channel is in one of the rate control
states. The power value will be displayed in inverse video.
MUST BE IN MANUAL
When in the Source Maintenance display the Toggle Sensor Shutter, Toggle
Source Shutter, and Crucible Switch functions can not be done unless the
Channel is in the Manual state.
MUST EDIT NAME FIRST
The user requested a save name or save message operation without first
having edited the name or message.
NO DISK IN DRIVE
Save or Load requires a floppy disk to be inserted: the disk is not detected by
the instrument.
NO GOOD CRYSTAL
An attempt is being made to Start a Channel when there is a crystal failure for
the Channel’s sensor.
NO ROTATE: CRYSTAL 2
A rotate crystal head function can not be done on a Crystal 2.
NO ROTATE: RUNNING
A rotate crystal head function can not be performed while the Channel process
is being executed.
NO SWITCH: BAD XTAL
An attempt is being made to switch the crystal and there are no good crystals
remaining in the sensor head. This message will only appear when the
Channel process is being executed. When in Ready a crystal switch to a failed
crystal is permitted.
NO SWITCH: NO BOARD
A crystal switch cannot be done because no measurement board is installed
for this channel.
NO SWITCH: NO OUTPUT
A crystal switch cannot be done because no output is set for crystal switching
in the Channel hardware parameters set up display.
NO SWITCH: SINGLE
A crystal switch cannot be requested while using a single head sensor.