Cygnus Operating Manual
NO SWITCH THIS DISP
The crystal switch function using the Hand Held controller is not allowed when
the instrument in on the current display.
NO TEST: NO BOARD
No XIU Test is performed if there is no measurement card for that channel.
NO TIME POWER AVG
Indicates a crystal failure occurred within the first minute of control when the
Averaging Time parameter value is >0, or a crystal failure occurred within the
first 2.5 seconds of control when the Averaging Time parameter value is =0.
NOT IN READY
Indicates the Channel cannot perform the requested function unless it is in the
READY state.
NOT ON CHANNEL
An attempt is made to do a START, STOP, RESET, or Crystal Switch from the
front panel when not on a display that selects a channel.
ONLY ONE CRUCIBLE
This message appears when a Crucible Switch function is attempted and the
Number of Crucibles value is 1.
OUTPUT USED EQU #
The output value being entered into the parameter is already in use by Logic
equation #.
PARAMETERS DEFAULTED
Indicates the instrument parameters have been set to their default values. Any
previously programmed values will have been changed to the default values.
POWER LOSS
Indicates the instrument lost line power immediately prior to the current power
on cycle.
POWER MUST BE ZERO
When in the Source Maintenance display the power must be zero before a
Crucible Switch function is performed.
PROGRAM LOCKED
Indicates the instrument is in Program Lock. This prohibits any parameter from
being entered from the front panel until the Program Lock Code is entered on
the Program display.
RATEWATCHER DELAY
The RateWatcher feature is active. The sensor is stabilizing for 5 seconds
before SAMPLE. The previous sample period’s Average Rate is being
integrated for thickness. The previous sample period’s Average Power is
being held constant.