Remote storage mill pwr, Operation manual, Common error messages – ACU-RITE MILLPWR Old Style User Manual

Page 126

background image

REMOTE STORAGE

MILLPWR

Operation Manual

Page 8-5

Common Error Messages

If the communication lines between

MILLPWR

and your PC are disrupted, an error

message will appear on the

MILLPWR

screen. Settings that do not match and a faulty

serial cable connection are the most common problems and are easiest to correct.

Error

Cause

Solution

BAUD rate
error

MILLPWR

and PC

BAUD rate settings do
not match.

Change one of the settings so that

MILLPWR

and the PC BAUD rates

are the same.

Checksum error

MILLPWR

and PC

BAUD rate settings do
not match.

Change one of the settings so that

MILLPWR

and the PC BAUD rates

are the same.

BAUD rate is too slow.

Increase the BAUD rates on both the

MILLPWR

and the PC.

No response
from PC

Serial cable connection
and COM port setting
do not match.

If the COM port setting (1 or 2) on the
PC screen does not correspond with
the port the serial cable is connected
into:

a) Change the setting; -OR-

b) Disconnect the cable from the

COM port that it is connected to
and connect it to a different COM
port.

Incorrect serial cable

Verify that you are using a serial cable
and that it is not a NULL modem
cable. Replace the cable if necessary.

MILLPWR

and PC

BAUD rate settings do
not match.

Change one of the settings so that the
BAUD rates for

MILLPWR

and your

PC are the same.