External power – ADS Environmental Services FlowShark QR 775003 A2 User Manual
Page 208
Maintenance and Troubleshooting 8-23
Problem
Data is missing at the beginning or end of the date
range following data collection.
Possible
Causes
Monitor activation may have failed.
Monitor time may be incorrect.
Monitor’s firmware file may be corrupt.
Possible
Solutions
Verify whether the monitor has been activated, and
activate it if necessary.
Check monitor time, and reset clock if necessary.
Run diagnostics in Profile to verify whether a firmware
problem may exist.
Contact your regional ADS representative.
Problem
An I/O error message displays when communicating
with the monitor.
Possible
Cause
Processor or analog board may be faulty.
Possible
Solutions
Re-attempt communication with the monitor.
Replace the monitor if defective.
Contact your regional ADS representative.
External Power
The following table contains troubleshooting techniques for
problems with the ADS External Power Supply. If another power
supply is in use, similar troubleshooting techniques may apply.
However, consult the manufacturer’s product documentation prior
to beginning the troubleshooting process.