beautypg.com

Replacing the desiccant – ADS Environmental Services FlowShark QR 775003 A2 User Manual

Page 198

background image

Maintenance and Troubleshooting 8-13

7. Evaluate the pressure depth sensor performance for a limited

time to ensure no damage has occurred to the sensor

electronics.

Replacing the Desiccant

When replacing the desiccant inside the pressure depth sensor dryer

tube, use indicating beads that are approximately 0.16 inches in

diameter. Smaller beads may be able to escape the dryer tube,

entering the clear plastic tubing leading to the sensor connector.

Desiccant beads of the appropriate size for this dryer tube are

available through ADS (p/n 901-11055-00).
Replace the desiccant in the following way:
1. Clip the cable ties securing the dryer tube to the monitor, when

applicable.

2. Use a 16-mm nut driver to remove the inset nut at the free end

of the dryer tube (end of the tube opposite the brass barbed

fitting). Do not use pliers to remove this nut from the tube!

Warning:

Do not attempt to access the desiccant in the

dryer tube through the end of the tube with the brass

fitting! This fitting is not designed for removal. Removing

the fitting will damage the dryer tube and break the

moisture seal on that end of the tube.

3. Use needle-nose pliers to gently remove the black foam

packing material just inside the opening.

4. Pour out the old desiccant beads from the dryer tube.
5. Refill an equivalent amount of new desiccant beads into the

tube.

6. Carefully replace the foam packing into the dryer tube.
7. Replace and tighten the nut on the dryer tube.
8. Secure the dryer tube to the monitor using cable ties, when

applicable.