8-22
ADS FlowShark IS Manual
Problem
Time on the monitor clock is incorrect.
Possible
Causes
Monitor clock may be a faulty.
PC clock may read incorrect time.
Possible
Solutions
Verify the time on the PC clock and correct if
necessary.
Reactivate the monitor to enable the clock.
Collect the data from the monitor and replace the
monitor if defective.
Problem
Profile provides a Device Time Out message.
Possible
Cause
Communications may be marginal.
Analog board may be faulty.
Possible
Solutions
Re-attempt communication with the monitor.
Replace the monitor if defective.
Contact your regional ADS representative.
Problem
Gap exists within the collected data.
Possible
Cause
Monitor time may be incorrect.
Monitor firmware file may be corrupt.
Possible
Solutions
Check monitor time, and reset clock if necessary.
Attempt to collect data within the gap.
Contact your regional ADS representative.