Editing the mli device – ADS Environmental Services FlowShark QR 775003 A2 User Manual
Page 148

Configuration and Activation 6-19
Access the ADS websit
Internet.
Select ADS Environmental Services > Support > Client
Services
.
Click on the Download pressure coefficients… link.
Right-click on the year in which the sensor was
manufactured, and select the Save Target As option.
Rename the file to reflect the following format:
pressure.xx
, where xx represents the last two digits of the
year in which the sensor was manufactured. Remove the
.txt extension!
Save the file to the appropriate location in the local file
system.
Pipe Height
This field displays the pipe height at the monitor
location. This parameter is not editable from this location.
Physical Offset
Enter the vertical distance from the top of
the pressure depth sensor to the bottom center of the pipe. ADS
recommends measuring this distance manually during the
monitor installation or confirmation process. Profile defaults to
0 inches
.
Electronic Offset
Enter the difference that exists between
the pressure depth sensor readings and the manually measured
depth. The default offset is 0 inches.
Editing the MLI Device
Typically, the MLI device does not require editing. However, if the
monitor will be supporting activities such as event notification or
dual data rates, refer to the Profile User’s Guide (#950015**) for
detailed instructions on properly setting the applicable parameters.