ADS FlowShark IS Manual
Upgrading the Monitor Firmware
Profile
enables you to download updated firmware to the
FlowShark IS flow monitor that may include new features and
capabilities or performance improvements and enhancements in
functions such as data processing, analysis, and communications.
ADS recommends collecting the data from the monitor before
updating the firmware in the monitor to avoid losing the flow data
stored in the monitor memory. Refer to the Profile User’s Guide
(#950015**) for information on collecting data from the monitor.
ADS also requires reactivating the monitor after updating the
firmware.
Upgrade the firmware in the monitor memory in the following way:
1. Select the monitor location for which you want to upgrade the
firmware, and select Tools > Diagnostics from the main menu
or
click on the Diagnostics toolbar button.
Diagnostics toolbar button
The Monitor Diagnostics dialog displays.