beautypg.com

Yokogawa DAQWORX User Manual

Page 66

background image

2-36

IM WX103-01E

• Handling of NaN values

• In the case of a TLOG function (tlogmax, tlogmin, tlogpp, tlogsum, and tlogave)

If NAN exists in the channel values, it is handled as though the value did not exist.

• In the case of the IsNan() function

This function can determine whether the value is NaN.

• In the case of min, max, sum, ave, and pp functions

NaN values are discarded.

• In the case of &&, ||, ^^, and ! operators

NaN values are considered true.

• For all other computations

Computation is performed taking NaN as a value. Thus, NaN is returned as the

computed result.

• The numeric value display on the Monitor screen displays INVALID.

• Handling when the value of a measurement channel is ±Over (exceeding the upper

and lower limits of the measurement range)

If the measured value used in the reference function or TLOG function is ±Over,

computation continues in one of two ways. Select the handling method using the

Calculate +Over/-Over as the MAX/MIN value of a range check box.

• Continue the computation as ±Over.

• Continue the calculation by setting ±Over to the maximum or minimum value of the

measurement range of the specified channel.
Use the following values as the maximum and minimum value of the measurement

range.

• DC voltage: ±10% of the range

However, 0/+63.000 mV for 60-mV (HQ) range and 0/+6.3000 V for 6-V (HQ)

range.

Example:

6 V range

+OVER: +6.6 V

–OVER: –6.6 V

• Temperature: ±10°C of the rated measurement range

However, 0 K and 10 K for KpVsAu7Fe and J263B.

• Linear scaling: Maximum and minimum values of the scale corresponding to

±10% of the measurement span

However, ±32000 when the scale value exceeds ±32000.

Example 1: When the measurement span is set to ±1 V and the scale is set to

±10000 at 2 V range

+OVER: +22000

–OVER: –22000

Example 2: When the measurement span is set to ±1 V and the scale is set to

±30000 at 2 V range

+OVER: +32000

–OVER: –32000

To suppress ±Over from occurring in the computation, use the limit function.

• Alarms for computation using the Alarm() function

The alarm corresponding to the current value is used for the alarm of a measurement

channel.

The alarm corresponding to the current value is also used for the alarm of a

computation channel. However, for computation channels that have not yet computed

the current value, the alarm corresponding to the previous value is used.

2.5 Settng Computatons (Settng the Computaton Channels)

This manual is related to the following products: