Yokogawa DAQWORX User Manual
Page 154

4-10
IM WX103-01E
4.2 Calbraton Procedure
• Six-Channel, Medium-Speed Four-Wire RTD Resistance Input Module (MX110-
V4R-M06)
Measurement Range to Be Calbrated
Input Value 1
Input Value 2
20 mV
0 mV on Ch1
20 mV on Ch2
60 mV
0 mV on Ch1
60 mV on Ch2
200 mV
0 mV on Ch1
200 mV on Ch2
1 V
0 V on Ch1
1 V on Ch2
2 V
0 V on Ch1
2 V on Ch2
6 V
0 V on Ch1
6 V on Ch2
20 V
0 V on Ch1
20 V on Ch2
100 V
0 V on Ch1
100 V on Ch2
RTD (1 mA) 20 mV
0 Ω on Ch3
20 Ω on Ch4
RTD (1 mA) 60 mV
0 Ω on Ch3
60 Ω on Ch5
RTD (1 mA) 200 mV
0 Ω on Ch3
200 Ω on Ch6
RTD (1 mA) 600 mV
0 Ω on Ch3
300 Ω on Ch4
RTD (0.25 mA) 600 mV
0 Ω on Ch3
2400 Ω on Ch5
RTD (0.25 mA) 1 V
0 Ω on Ch3
3000 Ω on Ch6
• 4-CH, Medium-Speed Strain Input Module (MX112-B12-M04, MX112-B35-M04,
MX112-NDI-M04)
Measurement Range to Be Calbrated
Connect1
Connect2
2000 µSTR (strain)
120.000 Ω on Ch2
117.154 Ω on Ch2
20000 µSTR (strain)
120.000 Ω on Ch2
113.010 Ω on Ch2
200000 µSTR (strain)
120.000 Ω on Ch2
80,000 Ω on Ch2
Use a 4-gauge method connection when performing calibration. For information on
this connection, see section 2.4 of the MX100 Data Acquisition Unit User’s Manual (IM
MX100-01E).
• 8-CH, Medium-Speed Analog Output Module (MX120-VAO-M08)
Adjust so that all channels output 0 V and 10 V. This differs from other input modules.
Range to Be Calbrated
Output val. 1
Output val. 2
10 V
0 V
10 V
Note
For a description of the measurement range and accuracy of each input module, see sections 4.4
and 4.5 in the MX100 Data Acquisition Unit User’s Manual (IM MX100-01E).