Yokogawa DAQWORX User Manual
Page 34

2-4
IM WX103-01E
Subnet Mask
Enter the mask value used when determining the subnet network address from the IP
address. The default value is 255.255.255.0. When connecting to the PC locally, this
value does not need to be changed. In this case, set the subnet mask on the PC also to
255.255.255.0.
Note
Consult your network administrator for the subnet mask value. You may not need to set the
value.
Default Gateway
Set the IP address of the gateway (default gateway) used to communicate with other
networks. The default value is 0.0.0.0. When connecting to the PC locally, this value
does not need to be changed. In this case, you do not have to set the default gateway on
the PC.
Note
Consult your network administrator for the IP address of the default gateway.
Connected User (User)
Displays the computer name of the PC that is connected to the MX100. This item is
blank (- - - - -) if there is no PC connected to it.
Software
Displays the name of the software program that is connected to the MX100. This item is
blank (- - - - -) if there is no software program connected to it.
• MXLOGGER:
Logger of the MXLOGGER
• MX100 Standard: Integration Monitor of the MX100 Standard Software
• MX Calibrator:
Calibrator of the MXLOGGER or the MX100 Standard Software
• API:
Software application created using the MX100/DARWIN API
(product sold separately)
2.1 Searchng MX100 Data Acquston Unts and Confgurng the Network